SILVER AUTOMATION INSTRUMENTS
English

Coriolis Flow Meter Accuracy: Grades, Calibration Reports and Density


Coriolis flow meter accuracy is usually stated as a percentage of the reading, such as ±0.2% for liquid mass flow, and every meter is checked against that figure before it leaves the factory. Across the industry, liquid mass flow accuracy for Coriolis meters typically falls between ±0.1% and ±0.5% of reading. For the Silver Automation Instruments SH-CMF coriolis mass flow meter, liquid accuracy grades of 0.1%, 0.15%, 0.2% and 0.5% are available, gas accuracy is ±0.5%, and each meter ships with a factory calibration report.

This guide helps engineers and buyers choose the right Coriolis meter accuracy grade, understand what a datasheet figure really means, and check a factory calibration report. It uses real SH-CMF calibration reports as examples.

coriolis flow meter accuracy calibration

Quick Answer

QuestionShort answer
What accuracy can a Coriolis meter reach?±0.1% to ±0.5% of reading for liquid mass flow, depending on the grade ordered; ±0.5% for gas (SH-CMF)
Which grade do most buyers choose?±0.2%. About three in four SH-CMF meters we supplied were ordered at this grade
Is accuracy the same across the whole range?No. Below roughly 10% of full scale, zero stability adds to the error
Does every meter come with a calibration report?Yes. Each meter is water-tested at 4 flow points before shipment
Is density in the calibration report?No. The standard report covers mass flow; density accuracy is a specification, typically ±0.001–0.002 g/cm³ for liquids
Can third-party calibration be arranged?Yes, at extra cost and with a longer delivery time

What Does Coriolis Flow Meter Accuracy Mean?

Coriolis flow meter accuracy is the largest difference allowed between the meter reading and the true mass flow, stated as a percentage of the actual reading. A meter rated ±0.2% that reads 1,000 kg should show a value between 998 and 1,002 kg when the true quantity is 1,000 kg.

Two related terms often appear next to the accuracy figure:

· Repeatability is how closely the meter repeats the same reading under the same conditions. A Coriolis meter's repeatability is usually several times better than its accuracy; the SH-CMF reports below show 0.03%.

· Zero stability is the small flow the meter may show when nothing is flowing. Zero stability is a fixed amount in kg/h, not a percentage, so the error it causes gets larger as the flow gets smaller.

"Of reading" (or "of rate") is different from "of full scale". A ±0.2% of reading meter at 10% of its range is still allowed only 0.2% of that small reading, while a ±0.2% of full scale meter would be allowed 2% of the reading at the same point. Coriolis meters are specified as a percentage of reading, which is one reason they suit wide flow ranges.

Which Accuracy Grade Should You Order?

The accuracy grade should match how the reading is used: for billing and high-value dosing, pay for a tighter grade; for process monitoring, ±0.2% or ±0.5% is normally enough.

Grade (liquid mass flow)Typical useShare of SH-CMF orders*
±0.1%Custody transfer, high-value products, reference metersAbout 3%
±0.15%Tighter batching where ±0.2% is borderlineAbout 2%
±0.2%Most process, batching and dosing linesAbout 76%
±0.5%Monitoring, utility lines, gas serviceAbout 19%

*Based on 105 SH-CMF order lines with a stated accuracy in our sales records.

The pattern in our quotations is worth noting. Many buyers ask for ±0.1% at the start, then choose ±0.2% once the price difference and the actual need are compared. Before asking for ±0.1%, check two things: whether the rest of the system (scale, tank gauge, lab analysis) is that accurate, and whether the meter will run mostly in the upper part of its range. A ±0.1% meter running at 5% of its range will not deliver ±0.1%.

SH CMF Coriolis mass flow meter nameplate

Why Does Accuracy Drop at Low Flow?

Accuracy drops at low flow because the zero stability error stays the same in kg/h while the flow gets smaller, so it becomes a larger share of the reading. Above roughly 10% of full scale, the stated grade applies; below it, the error follows a simple formula:

Error (%) = ± (zero stability ÷ flow rate) × 100

The zero stability figure differs by meter size and is listed in the datasheet. Two practical consequences follow:

  1. Size the meter for the real flow, not the pipe. A Coriolis meter is often one or two sizes smaller than the line. Running a large meter at a few percent of its range is the most common cause of disappointing accuracy.
  2. Zero the meter on site. Fill the meter with process fluid at operating temperature and pressure, stop the flow completely, then run the zero procedure. A good zero keeps the low-flow error close to the datasheet value.

For flows below about 10 kg/h, a dedicated low-flow meter is usually a better choice; the micro Coriolis selection guide covers that range.

coriolis low flow error zero stability curve

What Is in a Coriolis Calibration Report?

A Coriolis calibration report is the factory test record that shows how far the meter's reading was from the reference at several flow rates. Each Silver Automation Instruments SH-CMF meter is tested on water at our factory before shipment, and the report is sent with the meter or by email on request. The report has two pages.

Page 1 – Test report cover

FieldWhat it tells you
Report numberUnique number for this test
Model / specificationFull SH-CMF model code, including the accuracy code (A2 = 0.2%, A4 = 0.5%)
Factory numberSerial number of the sensor; matches the nameplate
Test basisJJG 1038-2008, the verification regulation for Coriolis mass flow meters used by the factory
Test conclusionThe accuracy grade the meter passed, e.g. "Accuracy Level ±0.2%"
Tester, inspector, test dateWho tested and checked the meter, and when
calibration report of coriolis flow meter DN10 1

Page 2 – Testing result

FieldWhat it tells you
Testing medium, temperature, pressureTest conditions; water at about 15 °C and 0.2 MPa
Flow range and instrument rangeThe range set for your order, e.g. 60–600 kg/h; not necessarily the full sensor range
Range percentage and tested valueThe 4 test points: full scale, then 50%, 20% and 10% of the set range
Average accuracy EiThe error at each test point, in %
Accuracy level EThe largest error found across all points
Repeatability ErHow closely repeated runs agree
Output typePulse factor, e.g. 60,000 pulses per kg
Precision curveThe Ei values plotted against the test points

Our calibration information page describes the test rigs and methods used at the factory.

calibration report of coriolis flow meter DN10

How to Read the Numbers on the Report

Reading the report means checking that every test point is inside the ordered grade, and then looking at how the error changes across the range. Two recent SH-CMF reports show how this works.

PointDN10 meter, ±0.2% grade, range 600 kg/h3" meter, ±0.5% grade, range 12 t/h
Full scale600 kg/h: −0.16%12 t/h: −0.09%
50%300 kg/h: −0.13%6 t/h: −0.06%
20%120 kg/h: −0.14%2.4 t/h: −0.12%
10%60 kg/h: −0.08%1.2 t/h: −0.17%
Accuracy level E−0.16%−0.17%
Repeatability Er0.03%0.03%

Three points to check on any report:

  1. Every Ei is inside the grade. For the DN10 meter, the largest error is 0.16%, inside ±0.2%. For the 3" meter, all points are well inside ±0.5%.
  2. The tested range matches your process. The 3" meter was tested at 1.2–12 t/h because that was the range set for the order. If your real flow is outside the tested range, ask for the test points to be set to your range before shipment.
  3. Repeatability is much smaller than the error. An Er of 0.03% means the meter gives the same answer each time; a small, steady offset is what calibration is designed to measure.

The serial number on the report should match the nameplate. Keep the report with the meter records; it is the baseline for any later recalibration.

Density Measurement Accuracy

Density measurement accuracy on the SH-CMF is typically ±0.001 to ±0.002 g/cm³ for liquids, within a density range of about 0.5–2.5 g/cm³. The meter measures density from the vibration frequency of the tubes, and the reading can be shown on the display or sent as a 4–20 mA output. Temperature measurement accuracy is ±1 °C.

The standard factory report does not include a density or temperature test; it covers mass flow only. If density is used for concentration, Brix or quality control, tell us at the order stage so the density requirement is reviewed before the meter is built. Questions on density outputs, gas service and other options are answered in the SH-CMF FAQ.

Factory Calibration vs Third-Party Calibration

Factory calibration is the standard water test done on every meter before shipment, while third-party calibration is an extra test at an independent laboratory, used when a contract, regulator or auditor requires it.

PointFactory calibrationThird-party calibration
Included as standardYesNo, extra cost
Delivery timeStandard (10–12 working days for standard configurations)Longer
Test mediumWaterDepends on the laboratory
When to chooseProcess control, batching, internal accountingCustody transfer, contracts or audits that name an independent laboratory

Water is the normal test medium because a Coriolis meter measures mass directly, so a mass calibration on water applies to other liquids. Recalibration is usually recommended every one to two years. Shipping a meter back to the factory overseas is costly, so most users recalibrate at a local independent facility and compare the result with the original report.

Need a Specific Accuracy or Certificate?

Send us these details and we will confirm the grade, sizing and documents you need:

  • Fluid, flow range (minimum, normal, maximum), temperature and pressure
  • Required accuracy and whether the reading is used for billing
  • Whether you need density, and the accuracy required
  • Any documents required: factory report only, or third-party calibration

Silver Automation Instruments replies within 24 hours with a recommended model code and quotation. Contact our engineers with your details.

FAQ

Does every Coriolis meter ship with a calibration report?
Every SH-CMF Coriolis meter is water-tested at 4 flow points before shipment and comes with a factory calibration report showing the error at each point, the repeatability and the accuracy grade passed. A copy can be requested later by quoting the factory number on the nameplate.
Is a ±0.1% meter worth it over ±0.2%?
A ±0.1% meter is worth it when the reading is used for billing or high-value products and the meter runs mostly above 20% of its range. For process control and most batching, ±0.2% is the common choice; about three in four of the SH-CMF meters we supplied were ordered at ±0.2%.
Can the meter be calibrated with oil instead of water?
The standard factory calibration uses water, which is normal for Coriolis meters because mass flow measurement does not depend on the fluid. If a contract requires calibration on a specific fluid, tell us at the enquiry stage so the options can be checked.
Is density included in the calibration report?
Density is not included in the standard calibration report, which covers mass flow only. Density accuracy for liquids is typically ±0.001 to ±0.002 g/cm³. If density is critical for your process, state the requirement when ordering.
How often should a Coriolis meter be recalibrated?
A Coriolis meter is usually recalibrated every one to two years, or as required by your quality system. Between calibrations, a zero check at operating conditions and a comparison against a weighed batch are simple ways to confirm the meter is still in tolerance.
Email
sales@silverinstruments.com
WA
WhatsApp QR Scan to WhatsApp
Inquiry
Send a Quote