Thermal Mass Flow Meters for Multi-Gas Measurement


Case Study: Five-Gas Thermal Mass Flow Meter Application — Hydrogen, Propane, Oxygen, Nitrogen, and Chlorine Application: Laboratory / Process Gas Control  |  Product: Silver SRK-DL Micro Thermal Mass Flow Meter |  Tags: MFC-140101 to MFC-140105

Quick Answer

Silver Automation Instruments supplied five SRK-DL micro thermal mass flow meters for a multi-gas laboratory application covering hydrogen, propane, oxygen, nitrogen, and chlorine. All five tags share a common 1/4" 3000# threaded connection, 24V DC / 4-20 mA / RS485 interface, and a 3 MPa pressure-rated body — selected as a cost-effective alternative to Western-brand MFC instruments with equivalent measurement performance.

Low flow gas Mass Flow Meter Installation

Project Background

A customer sent us an instrument data sheet covering five mass flow controllers on a single P&ID — B01-100-PR-DWG-14-002. Each tag runs a different gas: hydrogen, propane, oxygen, nitrogen, and chlorine. Flow rates are small, all in the range of 0.02 to 0.2 kg/h at normal conditions. Design pressure is 6 barg. Connection is 1/4" 3000# threaded throughout.

The original specification called for globe-type control valves with positioners (4-20 mA signal, M20x1.5 cable entry, equi-percentage characteristic). The customer was looking for a capable, budget-friendly supply source to cover the measurement portion of that scope. We proposed the SRK-DL series thermal mass flow meter — one platform, five gases, same wiring, same signal.

SRK-DL series thermal mass flow meters for multi-gas

Instrument Summary

TagFluidMWNormal FlowMax FlowPressureNotes
MFC-140101Hydrogen (H2)2.020.1 kg/h0.2 kg/h2 bara / 6 barg002L range, Fluororubber
MFC-140102Propane (C3H₈)44.10.1 kg/h0.2 kg/h2 bara / 6 barg500C range, Fluororubber
MFC-140103Oxygen (O2)320.1 kg/h0.2 kg/h2 bara / 6 barg200C range, O₂-clean
MFC-140104Nitrogen (N2)280.15 kg/h0.25 kg/h2 bara / 6 barg200C range, Fluororubber
MFC-140105Chlorine (Cl2)70.90.1 kg/h0.2 kg/h2 bara / 6 barg200C range, confirm wetted mat.

Tag-by-Tag Application Notes

MFC-140101 — Hydrogen Gas

Hydrogen is the lightest gas on the list, with a molecular weight of 2.02. That low density means the SLM-equivalent flow range is relatively large for a small mass flow. We specified the 2 SLM range (002L), which gives comfortable span coverage at the stated maximum of 0.2 kg/h. Fluororubber seals handle the application without issue at the 2 bara normal inlet pressure.

One thing to flag: the datasheet shows cryogenic design temperatures down to -253°C (near liquid hydrogen territory). The SRK-DL is rated 0 to 50°C at the meter body. We confirmed with the customer that the meter itself sits at ambient temperature — the extreme low temperature is a line design condition, not an operating condition at the instrument location.

MFC-140102 — Propane Gas (C3H8)

Propane has a molecular weight of 44.1, roughly 22 times heavier than hydrogen. The 500 SCCM range (500C) converts to approximately 75 SCCM equivalent for C3H8, keeping the normal flow well within the linear range. Fluororubber seals are compatible with dry propane gas service.

Propane is flammable. The customer's line class (S82A) and the 3000# threaded connection are appropriate for the 6 barg design pressure. If the installation zone requires ATEX or IECEx certification, the standard SRK-DL does not carry explosion-proof ratings — that would require a discussion on alternative enclosures.

MFC-140103 — Oxygen Gas (O2)

Oxygen service requires special attention. Organic contamination inside an oxygen line can ignite under pressure. We specified the 200 SCCM range for O2 (MW 32), which covers the 0.2 kg/h maximum. Fluororubber seals are generally compatible with gaseous O₂ at moderate pressures.

The key requirement here is O2-clean degreasing of all wetted parts before dispatch. This is a factory-confirmed process step. We flagged this explicitly in the quotation notes, and the customer must confirm it before placing the order. Skipping this step on an oxygen line is not acceptable practice.

MFC-140104 — Nitrogen Gas (N2)

Nitrogen is the most straightforward of the five gases — inert, dry, non-corrosive. MW is 28, and the 200 SCCM range comfortably covers the 0.25 kg/h maximum. Standard Fluororubber seals apply. No special treatment or confirmation required. This tag is a textbook thermal MFM application.

MFC-140105 — Chlorine Gas (Cl2)

Chlorine is the most aggressive fluid in this set. MW is 70.9, the heaviest gas here, and it is highly corrosive — especially in the presence of any moisture. The 200 SCCM range covers the 0.2 kg/h maximum at normal flow.

We raised a material compatibility question on this tag. SS316 wetted parts may be adequate for dry Cl2gas at low concentrations; we also use perfluoroelastomer (FFKM) as the sealing material to improve the corrosion resistance for chlorine measurement. The customer needs to confirm the Cl2 concentration and moisture content before we finalise the material selection.

SRK DL thermal mass flow meter for multi gas

SRK-DL thermal mass flow meter

The customer's original instruments were specified from a European brand. Lead times for that brand in this region run 10 to 16 weeks. The SRK-DL ships in 7 to 10 working days from China. For a project commissioning schedule, that difference matters.

Beyond lead time, there are several technical reasons the SRK-DL was a good fit here:

• One model platform covers all five gases. Same body, same wiring, same signal (4-20 mA + RS485). The instrument technician deals with one product, not five.

• 1/4" 3000# threaded connection matches the customer's line class exactly. No adapters.

• 3 MPa pressure rating — well above the 6 barg design pressure.

• RS485 output allows integration with a site SCADA or DCS without additional signal converters.

• Operating range 0 to 50°C covers the ambient conditions at the instrument location.

Western-brand thermal MFMs for this kind of multi-gas laboratory service typically run at a significant cost premium, often two to four times the per-unit cost of the SRK-DL. For five instruments of similar specification, the budget difference is material. The SRK-DL is not a stripped-down product — it offers the same 4-20 mA / RS485 dual output, integral display, and SS316 wetted construction that the European alternatives provide.

SRK-DL Technical Specifications (Common to All Five Tags)

ParameterValue
Measurement principleThermal mass flow (bypass capillary type)
Gas flow range2 SCCM to 30 SLM (model-dependent)
Accuracy±1.5% F.S. (typical)
Repeatability±0.5% F.S.
Pressure rating3 MPa (30 bar)
Operating temp.0 to 50°C
Output signal4-20 mA + RS485 (Modbus RTU)
Power supply24V DC
DisplayIntegral LCD, instant + totalised flow
Seal materialFluororubber (standard); O₂-clean on request
Process connection1/4" compression / G-thread (standard)
Body materialSS316
ProtectionIP65

Open Items Before Order

Three items need customer confirmation before this order can be placed:

• 1. MFC-140103 (O2): Confirm O2-clean degreasing requirement with factory. This is non-negotiable for oxygen service.

• 2. MFC-140105 (Cl2): Confirm Cl2 concentration and moisture content. If wet chlorine or high concentration, Hastelloy C-276 wetted parts are required instead of SS316.

• 3. Process connection: The SRK-DL ships with a 1/4" G-thread compression fitting as standard. If the customer's piping uses NPT threads or socket weld ends, confirm whether a thread adapter is acceptable or a special end connection is needed.


Request a Quote for Your Application

If you have a similar multi-gas flow measurement or control application, send us the following and we will size and quote within 24 hours:

  • Gas name and molecular weight (MW)
  • Normal and maximum mass flow rate (kg/h or SLPM)
  • Operating pressure (bar / bara / barg)
  • Operating temperature at the meter body (°C)
  •  Required output: 4-20 mA, RS485, or both
  • Process connection: thread type and size 

Contact Silver Automation Instruments: inquiry@silverinstruments.com  |  WhatsApp +86-18936759191  |  SRK-DL product page 


FAQ

Can one thermal mass flow meter model handle multiple gas types?
Yes, Each SRK-DL unit is calibrated by N2, for its specific gas, we will do gas coefficient adjustment. The hardware is identical across all five tags in this project. What changes is the internal calibration table and the flow range selection, which is set at the factory based on the gas molecular weight and the target flow range. You cannot field-switch between gases without recalibration.
What is the difference between a thermal mass flow meter and a thermal mass flow controller?
A flow meter (MFM) measures flow and outputs a signal — 4-20 mA, RS485, or both. A flow controller (MFC) adds an integrated control valve to regulate flow to a setpoint. This project specified the measurement-only (MFM) version. The SRK-DL is available as both; the MFC variant includes an internal proportional valve and costs more.
Is the SRK-DL suitable for corrosive gases like chlorine?
Standard SS316 wetted parts may be adequate for dry chlorine gas at low concentrations. For wet Cl2 or higher concentrations, Hastelloy C-276 wetted parts are the correct choice. We flag this for every Cl2 inquiry and will not supply SS316 without explicit confirmation from the customer that the service conditions are within acceptable limits.
What certifications does the SRK-DL carry?
The standard SRK-DL carries CE marking. It does not carry ATEX or IECEx explosion-proof certification in its standard configuration.
How does Silver Instruments compare on lead time versus European MFC suppliers?
Standard SRK-DL units ship in 7 to 10 working days from our China facility. European-brand equivalents for this product class commonly carry 10 to 16 week lead times when ordered through local distributors in Asia. For multi-tag instrument packages on a project schedule, that gap has a direct impact on commissioning timelines.