SILVER AUTOMATION INSTRUMENTS
English

Micro Low-Flow Coriolis Mass Flow Meter Selection Guide


A micro low-flow Coriolis mass flow meter is sized mainly from the required mass-flow range, not simply from the process pipe diameter. A practical starting point is to keep normal flow around 50% to 80% of full scale, while also allowing for pressure drop, wetted materials, and hazardous-area requirements.

The examples below reflect typical requirements for refrigerant skids, dosing benches, and calibration labs, using the process conditions provided for each application.

TEX approved micro mass flow meter for refrigerant measurement

1. What Counts as Micro or Low Flow

There is no universal threshold for micro or low flow. In practice, micro Coriolis meters cover flows from a few grams per minute to a few hundred kg/h, with sensor bores of about DN1.5 to DN15 (1/16 in to 1/2 in).

Typical flows from our inquiries include:

ApplicationTypical flow
Peroxide or additive dosing bench90 to 105 mL/min
Calibration laboratory reference100 to 5,000 mL/min
Refrigerant liquid line (heat pump)30 to 150 kg/h
Pilot or skid-mounted process line5 to 500 kg/h

If the required flow is below this range, a different flow meter technology may be a better fit. Please see Section 9 for the main alternatives.

2. Convert to Mass Flow First

A Coriolis meter measures mass flow, while the process data may be given in L/min or mL/min. Convert the volumetric flow using the fluid density at the operating temperature before sizing the meter.

Water shows why temperature matters. At 5,000 mL/min:

  • At 20 °C (0.998 g/mL): about 299 kg/h
  • At 60 °C (0.983 g/mL): about 295 kg/h

The difference is about 1.5%, which is larger than the accuracy of the meter in this example. Refrigerants and most organic solvents can show a greater density change than water.

The pipeline diameter alone is therefore not enough to size a Coriolis meter.

3. Size by Flow Range, Not Pipe Diameter

Pipe size is a common starting point in a flow-meter inquiry, but it does not determine the correct Coriolis sensor bore. Two examples show why.

Inquiry example A: refrigerant line

  • Application: liquid refrigerant line in a heat pump
  • Customer data: φ10 mm copper tube
  • Recommended: DN1.5 to DN3 sensor

The sensor size should be based on the required flow range rather than the tube diameter alone. A DN10 sensor may also be suitable if the operating flow falls within an appropriate part of its measurement range.

Inquiry example B: calibration bench

  • Application: calibration of water volumes
  • Customer data: up to 5,000 mL/min (about 300 kg/h)
  • Recommended: around DN6, depending on the meter's flow range and pressure drop

Here, the sensor bore follows the required flow range rather than the line size.

As a practical starting point, aim for normal flow around 50% to 80% of full scale. Higher utilization can increase pressure drop, so the maximum-flow condition also needs attention.

SH CMF FE series small flow coriolis flow controller

4. Accuracy, Zero Stability and Turndown

At low flow, two accuracy terms matter: percent of reading and zero stability. Percent-of-reading error changes with the measured flow, while zero stability adds a fixed error in kg/h. As the flow decreases, that fixed term becomes a larger share of the reading.

For example, with a zero stability of 0.01 kg/h, the zero term is 0.1% at 10 kg/h but 1% at 1 kg/h. The meter has not changed, but the same zero error now represents ten times the percentage of the measured flow.

For percent-of-reading accuracy, ±0.2% or better is a realistic target for many micro meters. The more useful specification is the zero stability and the accuracy statement at the actual operating flow, rather than the headline accuracy alone.

Turndown matters for the same reason. A sensor selected for 30 to 150 kg/h (5:1) keeps the normal operating point in a more useful part of the range than a sensor stretched across 1:100.

5. Pressure Drop vs Pump Head

Small-bore Coriolis sensors can create noticeable pressure drop (Δp), and low-flow skids often have limited pump head. Compare Δp at normal and maximum flow with the available pump head before finalizing the meter.

If pressure drop is the limiting factor, a slightly larger bore can be a better trade-off, provided normal flow still sits in a suitable part of the meter range.

6. Wetted Materials and Fluid

The fluid affects the wetted material, seal selection, and in some cases the meter model. The fluid, concentration, and operating temperature should be stated together because each can change the material requirement and price.

  • Peroxides and aggressive chemicals: wetted materials depend on concentration and temperature; non-metallic or 316L stainless options may be considered where compatible.
  • Food, dairy, and pharma: hygienic finish and sanitary connections are relevant.
  • Fuels, solvents, and kerosene: 316L is commonly used, subject to the specific fluid and operating conditions.
  • Refrigerants: 316L with a sealed, dry transmitter is used for the stated applications.

7. Pressure, Vacuum and Hazardous Area

Micro Coriolis applications can cover a wide pressure envelope, especially on refrigerant skids. Two inquiries illustrate the point.

Inquiry example C: refrigerant skid. The line ran from 0.4 mbar abs (300 µmHg, evacuation) up to about 41 bar. The full pressure envelope needs to be stated because it affects construction and pressure rating.

Inquiry example D: heat pump unit. The customer required ATEX approval for A2L refrigerants (R454C, R454B, R32) and a remote display because the meter sits inside a compact unit.

Certificate type, temperature class, cable length, and enclosure rating should be defined early. These configuration details can affect both price and lead time in micro-flow applications.

8. Installation Traps for Micro Meters

Because of their small internal bore and low flow rates, micro flow meters require careful attention to installation conditions:

  • Vibration: Micro sensors with a single-tube loop can react more strongly to external vibration than dual-tube designs. Rigid support and some distance from pumps and compressors can help keep the signal stable.
  • Pump pulsation: Diaphragm and peristaltic pumps can create pulses. Transmitter damping or a pulsation dampener may be needed when the pulsation is significant.
  • Bubbles and flashing: The pipe should remain full. On refrigerant lines, install the meter where the fluid is fully liquid, such as after the condenser or subcooler, and keep control valves downstream where practical.
  • Debris: The bore is small, so a suitable upstream strainer can help protect the sensor.
  • Zero adjustment: Set zero at true zero flow, with a full pipe and the meter at operating temperature.

9. When Micro Coriolis Is the Wrong Choice

Coriolis offers high measurement accuracy, little straight-pipe requirement, and direct mass-flow measurement. These advantages matter when fluid properties change or the application needs tight accuracy. If the application does not need them, a turbine or oval gear meter may provide the required measurement at lower cost.

A Coriolis meter typically costs several times more than a turbine or oval gear meter.

TechnologyFluidsBest fit
Micro CoriolisLiquid and gasMass flow, changing fluid properties, tight accuracy
Micro oval gearLiquid, including viscousViscous liquid, dosing, cost-sensitive builds
Low-flow turbineClean, non-corrosive liquidWater, light oil, volumetric calibration
Thermal massGas onlyLow-flow gas where Coriolis is not economical

For a wider technology overview, see our low-flow flow meter guide.

10. What to Confirm Before Sizing

The following information is enough to establish the main sizing and configuration requirements:

  • Fluid, density, viscosity, and temperature affect wetted material and calibration requirements.
  • Normal and maximum flow, preferably in mass units, determine the sensor bore and useful turndown.
  • Line size and process connection define the mechanical connection, such as 1/4 in NPT, BSP, or a compression fitting for 10 mm OD tube.
  • Pressure range, including vacuum, affects the required construction and pressure rating.
  • Allowable pressure drop matters when the available pump head is limited, especially at maximum flow.
  • Accuracy should be considered at the actual operating flow rather than only from the headline specification.
  • For hazardous-area installations or A2L refrigerants, state the required certification, temperature class, and installation conditions.
  • Outputs, display, and totalizer functions are configuration items. State requirements such as 4-20 mA, Modbus RS485, remote display, or a per-minute reading at the inquiry stage.

FAQ of Micro Coriolis Meters

What is the smallest Coriolis flow meter size?
Micro models are available down to around DN1.5, with full ranges in the grams-to-kilograms-per-hour class. Bore alone does not set the low end. Zero stability also matters, so compare it with the normal operating flow.
How accurate is a micro Coriolis meter at low flow?
Headline accuracy applies above a certain flow. At lower flow, the fixed zero-stability error becomes a larger share of the reading. A suitable sensor bore keeps normal flow in a useful part of the range.
Coriolis vs thermal mass meter for low-flow gas?
Thermal mass is suitable for very low gas flow and simple gases. Coriolis is useful when direct mass flow and density are needed, or when gas composition changes, provided the pressure is high enough to give usable gas density.
Can a micro Coriolis meter measure refrigerant liquid?
Yes, provided the refrigerant remains fully liquid at the meter. Install the meter after the condenser or subcooler and keep Δp low enough to avoid flashing. State the pressure envelope, including vacuum, and any required A2L approval.
When is an oval gear or turbine meter enough?
When volume measurement is sufficient, accuracy around 0.5% to 1% is acceptable, and the fluid is stable. Oval gear meters suit viscous liquids, while turbine meters suit clean, low-viscosity liquids.

Need Help Sizing Your Meter?

Send us these items and we can recommend a model and confirm the pressure drop:

  • Fluid, concentration, and temperature
  • Normal and maximum flow
  • Line size and process connection
  • Pressure range, including vacuum
  • Allowable pressure drop
  • Hazardous-area requirement

For available bores and flow ranges, see our micro Coriolis meter range.

Email sales@silverinstruments.com or WhatsApp +86 189 3675 9191.

Email
sales@silverinstruments.com
WA
WhatsApp QR Scan to WhatsApp
Inquiry
Send a Quote